Attention! The catalysis cell is currently unavailable for users. The high-temperature cell offers the same capabilities (and a lot more).
The Catalysis / general purpose AP cell enables photoemission measurements on catalytic and surface science type systems at mbar pressures. For the time-resolved APXPS please check Time-resolved APXPS page
Description | Small volume (1 L) cell with controllable gas supply. Standard setup for general users. |
Base Pressure | 2 x 10-9 mbar |
Sample manipulation | X, Y, Z Normal Emission. 55 deg synchrotron beam incidence (beam to surf normal) |
Pressure range | 2 x 10-9 – 30 mbar |
Heaters | Direct laser (CW, 808 nm) Up to 600C in 10 mbar for all gases except H2 and He. for H2 and He: – up to 600 C in 3 mbar – up to 500 C in 5 mbar – up to 400 C in 10 mbar PID controlled |
Cooling | Water cooling down to 17 C |
Sample holder type | Omicron flag style |
Sample size | Up to 8×8 mm2 |
QMS | Hiden HAL 3F/PIC Probe of inlet and outlet gas composition |
Analyser apertures | 0.3 mm |
Gas dosing (available 2018-I) | up to 8 individual gases (by user request) and their mixture. Fully automatic gas dosing system |
Max flow rate | 30 sccm (one gas) |
Manipulation | X, Y, Z. NE, 55 deg incidence (to surf. normal) |
Availability | unavailable for general users |